NITRATE-SELECTIVE ELECTRODES BASED ON THE TRINUCLEAR CHROMIUM(III) PIVALATES

Show simple item record

dc.contributor.author Revenco, Mihail
dc.contributor.author Martin, Mariana
dc.contributor.author Dayyih, Waell A.A.
dc.date.accessioned 2018-03-26T12:32:36Z
dc.date.available 2018-03-26T12:32:36Z
dc.date.issued 2008
dc.identifier.citation REVENCO, M., MARTIN, M. et al. Nitrate-selective electrodes based on the trinuclear chromium(III) pivalates. In: Cemistry Journal of Moldova. 2008, nr.1, pp. 44-47. ISSN 1857-1727. en
dc.identifier.issn 1857-1727
dc.identifier.uri http://dspace.usm.md:8080/xmlui/handle/123456789/1714
dc.description.abstract The paper describes the analytical potentialities of the trinuclear chromium(III) complexes as potentiometric ionophores for the construction of electrodes sensitive to the presence of nitrate anion. The electroactive material containing 4,4’-bipyridil was synthesized in situ. The membrane was prepared using dioctylphthalate as a solvent mediator and poly (vinyl chloride) as a polymeric matrix. The electrodes presented a slope of 56 mV/decade, a low limit of detection (3,2.10-6 mol/l), an adequate lifetime (4 months), and suitable selectivity characteristics when compared with other nitrate electrodes. The good parameters of this electrode made possible its application to the determination of nitrate in different types of fertilizers. en
dc.language.iso en en
dc.publisher Academia de Știinițe a Moldovei en
dc.subject trinuclearchromium(III) pivalate en
dc.subject nitrate-sensors en
dc.title NITRATE-SELECTIVE ELECTRODES BASED ON THE TRINUCLEAR CHROMIUM(III) PIVALATES en